The Valleylab Force FX Electrosurgical Generator is an isolated output electrosurgical generator that provides the power for cutting, desiccating, and fulgurating tissue during surgery. The generator is used for bipolar and monopolar electrosurgical procedures. You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated. The Force FX allows you to adjust the activation tone volume and reset the mode and power settings used previously. An RF activation port, RS-232 serial port, and expansion port are also included. The Valleylab Force FX Electrosurgical Generator can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor.
Valleylab Force FX Electrosurgical Generator Features
- Instant Response technology ensures that the power delivered remains virtually constant, regardless of the tissue type
- Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
- Three internal micro-controllers reduce system reaction time and increase the system’s processing speed
- Exclusive electrosurgical generator for the Computer Motion Hermes Voice Command System
- Spray coagulation voltage of no more than 9000 VAC peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
- A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
- (3) Cut Modes, all controlled by Instant Response technology, offer surgeons a variety of choices
- Low Cut for delicate tissue or laparoscopic cases
- Pure Cut for a clean, precise cut
- Blend for cutting with hemostasis
- (4) Coag Modes
- Desiccate for low voltage contact coagulation suitable for laparoscopic and delicate tissue work
- Fulgurate (high crest factor) for efficient non-contact coagulation in most applications
- Fulgurate (low crest factor) for lower voltage coagulation requirements
- Spray for coagulating large tissue areas with a superficial depth of necrosis
- (3) Bipolar Modes – controlled by the Instant Responseâ„¢ system
- Precise – utilizes low voltage to prevent sparking
- Standard – utilizes low voltage to prevent sparking
- Macro-bipolar are
- Compatible with and used as the electrosurgical energy source for
- Dyonics Control RF arthroscopic ablation system
- Dyonics Electrobladeâ„¢ rotary resection system
- Cook Vascular Perfectaâ„¢ EDS pacemaker lead extraction system
- Uniquely compatible with other devices
- Force Argonâ„¢ II and Force GSUâ„¢ argon coagulation system
- CUSA Excelâ„¢ and CUSAâ„¢ 200 ultrasonic surgical aspirators
- OptiMummâ„¢ smoke evacuator, through a direct cable link
- Valleylab VLCM bipolar current monitor
Valleylab Force FX Electrosurgical Generator Specifications
- BIPOLAR OUTPUT WAVEFORMS
- Precise: 470 kHz sinusoid
- Standard: 470 kHz sinusoid
- Macro: 470 kHz sinusoid
- MONOPOLAR CUT OUTPUT WAVEFORMS
- Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value
- Pure: 390 kHz sinusoid
- Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope
- MONOPOLAR COAG OUTPUT WAVEFORMS
- Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle
- Fulgurate: 390 kHz damped sinusoidal bursts with a repetition frequency of 30 or 57 kHz into 500 ohms
- Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21kHz < f < 35 kHz
- Input Power
- CUT: 85 – 132 VAC | 7 A
- COAG: 85 – 132 VAC | 4 A
- High Frequency Leakage
- Bipolar: less than 60 mA
- Monopolar: less than 150 mA
- Dimensions: 14″ W X 17″ D x 4 1/2″ H
- Weight: 18 lbs